52
Has the membrane
cap been attached
properly?
Attach the membrane cap
properly. If the membrane cap is
loose, the standard solution or
sample solution may enter inside
the electrode resulting in a
measurement error.
(
(page 8))
―
Has the glass
electrode for pH
measurement been
stored in the dry
state?
If it has been stored in the dry
state, it may cause measurement
errors or decrease in the
response rate during calibration.
Perform conditioning before
calibration.
(
Note on page 21)
If it will not be used for one week
or longer, store it according to the
storage procedure.
(
―
Are there air bubbles
inside the membrane
cap?
If there are air bubbles between
the membrane and the pH glass
ion selective membrane, it may
cause measurement errors or
calibration errors. Please check
the assembly procedure.
(
―
Check item
Solution
View in video