Maintenance
Membrane cap replacement
Replace the membrane cap as described in the following procedure.
1
Turn the electrode with the old
membrane cap upward and
remove the old membrane cap by
turning it counterclockwise.
2
Drain the internal solution by
turning the electrode with the
membrane cap downward and
opening the rubber stopper of
the internal solution fill port.
When disposing of the old membrane cap, follow the local rules
and regulations.
Tip
Tip