24
14
Shake the electrode to remove any air bubbles out of the
membrane.
15
Perform the second calibration.
Wait for a stable reading and perform the second calibration as described
in the Meter Instruction Manual.
When the second calibration is finished, the slope (%) of the electrode is
displayed.
16
Check the slope (%) on the meter.
For details on how to check the slope (%), refer to the Meter Instruction
Manual.
If air bubbles remain on the
membrane, it may result in inaccurate
measurement.
Do not hit the beaker with the
electrode. It may result in inaccurate
measuring.
Shake the
electrode
View video
View video
Note
Note