40
5
Wipe off the moisture.
Wipe off the moisture using soft and clean lint-free paper or cloth.
6
Make sure that there are no dirt or crystals of internal
solution adhered to the area around the glass electrode for
pH measurement and support pipe.
If dirt remains, repeat steps 4 and 5.
7
Fill 6 mL of the internal solution with the syringe into the
protective cap.
8
Attach the protective cap.
Store the glass electrode for pH measurement and the
membrane cap safely without any damage.
Do not get water and foreign matters inside of the protective
cap when attach it to the electrode. It may shorten the
lifespan.
When you will use the electrode again, perform operation
from filling with internal solution (
Note
Note