22
8
Shake the electrode to remove any air bubbles remaining
on the membrane.
9
Perform the first calibration.
Wait for a stable reading and perform the first calibration as described in
the Meter Instruction Manual.
10
When the first calibration is finished, take out the electrode
from the solution for calibration.
If air bubbles remain on the
membrane, it may result in inaccurate
measurement.
Do not hit the beaker with the
electrode. It may result in inaccurate
measuring.
Shake the
electrode
View video
View video
Note
Note