![horiba 5002S-10C Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/horiba/5002s-10c/5002s-10c_instruction-manual_2162179050.webp)
38
S
torage
When the electrode will not be used for one week or longer, store it as
described in the following procedure.
Main unit of electrode
1
Turn the electrode with the
membrane cap upward and
remove the membrane cap by
turning it counterclockwise.
2
Drain the internal solution by
turning the electrode with the
membrane cap downward and
opening the rubber stopper of
the internal solution fill port.