31
Electrode Preparation
Calibration
Measurement
After Measurement
M
easurement operation
Before operation
The calibration characteristics vary depending on the
temperature. The standard solution in calibration and the
sample solution in measurement shall have almost the same
temperature. Match the temperature as close as possible by,
for example, using a thermostatic bath. The temperature
difference may cause measurement errors.
Measure the sample solution only once after adding the pH
-
adjusting ISA.
Also, perform the measurement immediately after adding the
pH
-
adjusting ISA.
The ammonia in the stirred sample solution gets diffused as
time passes, causing the measurement result to vary.
* :
Prepare 5 mol/L aqueous solution of sodium hydroxide, or pur-
chase the pH-adjusting ISA available as consumables.
(
Note
Note
Syringe for
adjusting pH
pH-adjusting ISA
*
Items to be prepared