Electrode Preparation
Calibration
Measurement
After Measurement
M
easurement hints
A gradual rise in solution temperature due to the magnetic stirrer generating
heat may cause an apparent potential drift.
When it happens, reduce the influence of the temperature changes by
covering the magnetic stirrer with heat insulation such as polystyrene foam or
using a thermostatic bath.
Soak the membrane cap fully into the sample solution during measurement.
If the pH of the sample solution has exceeded the measurable pH range,
accurate measurement may not be possible (