DeltaMyc Operation Manual J810018 (7 Oct 2014)
Maintenance & Troubleshooting
4-4
The trinocular is set
for viewing the sam-
ple through the binoc-
ulars.
Slide the trinocular selection mir-
ror to another position (either
camera view or binocular/camera
view). Refer to the Olympus doc-
umentation if any doubt remains.
In the
Select point of
interest
window, the
stage does not move to
the selected point, when
you click on the point of
interest
The current objective
does not match the
objective selected in
DataStation.
Select the appropriate objective in
DataStation.
The hardware config-
uration has changed
(camera and stage
have been moved or
replaced).
Run the calibration procedure.
If there is no move-
ment at all, the stage
may be not operation-
al.
Make sure that the Tango control-
ler is ON and check the XY ca-
bles.
The
Start
or
Stop
ratemeter value is zero
The hardware is not
ON.
Check that the laser controller is
ON.
A signal cable is miss-
ing
Check the signal cable between
the DeltaDiode controller and the
DeltaHub. Check the signal cable
between the PPD detector and the
DeltaHub.
The
Start
or
Stop
ratemeter value is very
low
The ND filter value is
too high.
Select a lower ND filter value.
The pinhole is closed.
Check that the correct pinhole is
inserted properly.
The improper filter-
cube is selected.
Choose a correct filter-cube.
The microscope is in
“Imaging” mode.
Change the microscope mode to
“Time-resolved” in DataStation.