Troubleshooting
B.2 Fault
Description
Troubleshooting
Fault 1
Internal communication failure
Power-cycle the unit. If this fault occurs again, contact manufacturer.
Fault 2
Cell failure
Check the sensor connection. If this fault occurs again, replace sensor.
Fault 3
Cell is producing a negative reading
Calibrate the unit
Fault 4
EEPROM is corrupted
Reconfigure the unit by using mobile app. If this fault occurs again,
contact manufacturer.
Fault 5
MCU operating voltage failure
Power-cycle the unit. If this fault occurs again, contact manufacturer.
Fault 6
RAM read/write fault
Power-cycle the unit. If this fault occurs again, contact manufacturer.
Fault 7
Flash memory corrupted
Reconfigure the unit by using mobile app
Fault 8
Code memory failure
Power-cycle the unit. If this fault occurs again, contact manufacturer.
Fault 9
mA output failure (mA output version only)
Check that the input voltage is within specification.
Fault 10 Supplied voltage failure
Power-cycle the unit. If this fault occurs again, contact manufacturer.
Fault 11 Internal HW Fault
Power-cycle the unit. If this fault occurs again, contact manufacturer.
Fault 12 Internal SW Fault
Power-cycle the unit. If this fault occurs again, contact manufacturer.
Fault 13 Calibration Overdue
Calibrate the unit.