Modbus
Registers
Modbus Register Address Information
R/W
Type
Size Description
Notes
40116
Reserved
R
u16
1
40117
Temperature (Degrees C)
R
s16
1
Transmitter temperature
40118 to 40122
Reserved
40123 to 40131
Date and Time
R
string[18]
9
Format is “mm/dd/yy hh:mm:ss”
40132
Sensor Type
R
u8
1
0x00 :ECC
0x01: FL CAT
0x02: IR
0x03: PID
0x04: MOS
40133 to 40136
Reserved
40137 to 40138
Microcontroller Operating Voltage
R
f32
2
40139 to 40140
Microcontroller Reference Voltage
R
f32
2
40141
BLE Connection Status
R
u8
1
0x00: Not connected, 0x01: Connected
40142 to 40145
Reserved
40146 to 40152
Gas Name String
R
string[14]
7
40153
Temperature (Degrees C)
R
s16
1
Transmitter temperature, duplicate
40154
Temperature (Degrees F)
R
s16
1
40155
Calibration Due
R
u16
1
Calibration interval - DaysPassedSinceLastSuccessfulCal
40156
Relay Status
R
u8
1
High nibble : Relay 2 status
Low nibble : Relay 1 status
1 : Energized, 0 : De-energized
40157
Supply Voltage
R
f32
2
The voltage supplied to the Sensepoint XCL/XRL at the nominal 24.0
volt input, in millivolts.
40158
Reserved
40159
Calibration Interval
R
u16
1
40160
Alarm Type
R
u8
1
High nibble : Alarm 2 Type
Low nibble : Alarm 1 Type
0: Disable, 1: Rising, 2:Falling
40161
Inhibit Timeout
R
u16
1
Not used in Sensepoint XCL/XRL
40162
Relay Configuration
R
u8
1
1011
Bit: 0 to 1: Relay Type 1 (0: Fault, 1: Alarm1, 2: Alarm2)
Bit: 2 to 3: Relay Type 2 (0:Fault, 1: Alarm1, 2: Alarm2)
Bit: 4 : Initial Relay Status 1 (0: De-energized, 1: Energized)
Bit: 5 : Initial Relay Status 2 (0: De-energized, 1: Energized)
59