Maintenance
status and second flash indicates span status . The indicator flashes
long yellow
followed by short green flash
and short yellow
flash
when the zero calibration is successfully completed, but
flashes long yellow
followed by short red flash
and short yellow
flash
if it fails.
Span calibration
During span calibration, the indicator flashes yellow
. The indicator
flashes green
with two short pauses when the span calibration is
successfully completed, but flashes red
with two short pauses if it
fails.
Calibration gas purge
While the calibration gas is purged, the indicator is on steady yellow
.
Cancel
When a task is cancelled, the indicator flashes yellow
with three
short pauses.
4.3 Calibration Cap
The following table shows the recommended flow rate and stabilization
time for each of the gas types.
Gas Type
Recommended Flow Rate
Recommended Stabilization Time
O
2
500 ml/min
5 min
CH
4
500 ml/min
5 min
CO
500 ml/min
5 min
H
2
S
500 ml/min
5 min
H
2
500 ml/min
6 min
NOTE
Always use the calibration cap, tubing and fittings appropriate for
the type of gas being applied.
Note that the calibration cap, tubing and fittings are not part of the
explosion-proof certificate.