9 Software »Motion Manager« (optional)
Operating manual mzr-2542 / mzr
-
2942
58
Technical data subject to change without prior notice!
Last update: January 2016
The entry of command lines takes place in the field »Enter command:«. When
the command entry is completed this must be confirmed by pressing the button
»Send« to send. The commands can alternatively be written in capital or small
letters. Surplus blanks are ignored by the motion controller.
Example for feeding
Command Description
SOR0
V1000
V6000
V0
SOR1
Source For Velocity - Set command velocity with RS-232
Switch pump to velocity control mode and rotate with 1000 RPM
(equivalent flow for mzr-2942 = 3 ml/min)
Switch pump to velocity control mode and rotate with 6000 RPM
(equivalent flow for mzr-2942 = 18 ml/min)
Stop all pumps (rotate with 0 RPM)
Source For Velocity - Set command velocity with potentiometer
on the connection panel
Example for dosage
Command Description
SOR0
LR1280
M
LR2560
M
SOR1
Source For Velocity - Set command velocity with RS-232
Load relative position at 1280 to all pumps (1280 = 10
revolution, equivalent quantity mzr-2942 = 30
μ
l)
Start positioning for all pump
Load relative position at 2560 to all pumps (2560 = 20
revolutions, equivalent quantity mzr-2942 = 60
μ
l)
Start positioning for all pump
Source For Velocity - Set command velocity with potentiometer
on the connection panel
Because of the 4-edge evaluation of the encoder signal by the control unit
S-ND a quadruple number of counts per turn must always be given. In case
gear box module is used, the respective gear reduction must be taken into
consideration. For detailed information as to the use of Motion Manager,
please read the program online help.