Mount the parts of the bottle as on figure and insert the filling
hose in till the inside bottom of the bottle.
And set the burette and Beaker in front of the door of main box.
Depending of temperature conditions, you can also let the bottle
in front door of the main box, or set in a cooling system.
Don’t forget to connect the Serial link cable between terminal box
and computer. Finally connect the main power cable to the
equipment.
Starting on First Installation
The basic software is installed by using the
application. If this is done for the
first time, different operations should be done, to assure the correct installation of the
hardware machine parameters and measurement system.
1. Choose the
burette type
the terminal is connected to. Remember that not all kind of
burette types can be used with all DADS terminal versions. Check the DADS terminal
version overview on DADS Terminal Overview paragraph, to determine what type of
terminal supports which burette;
2. Fill the burette doing several Burette Purge till its full;
3. Verify the
burette debit
. See how to do on Burette Debit paragraph;
4. Make a
torque calibration
(for
type H
machine only). You can see how to do it on
Torque Calibration paragraph;
5. Define and choose the right
chamber
. The chamber can be defined in the HITEC
Config application and see its manual at
is chosen in the DADS application, which you can see on Menu Configuration /
Chamber;
6. In case the chamber is brand new a
Pre Polishing
has to be done, see how to do it
on Menu Maintenance / Pre-polish. This can be done during the night.
7. Make at least two
TLS tests
, in case you are still working with the Proc. A method.
You can see how to do it on Menu Maintenance / TLS.
8.
You have to make
the
Normalization tests
, which you can see how to do it on
Menu Maintenance / Normalization.
9. Generate
Quality files
, with the Quality Editor. You can see how to generate a
.
If all this is done, normal absorption tests can be done. Yo
u can see on “How to do a test?“.