Here below you have all the procedure of the table showing each step.
A
B
C
D
E
F
G
H
After you have written down the values for each weight step, go in the Action Log and write
in the measured factors cells, the
Torque Calibration
values.
If the R
2
-Value (correlation value) is 1.000000, the linearity of the strain gauge is proved. You
can repeat it twice. If you do not have approximately the same results, then you need maybe
to change the strain gauge and restart the manual service procedure.
10Kg