7.
End of test
(determined by
the
application)
"Q:xxx"
The application stops the test. One of the
test
results
is shown. The user can check the other
test results by pushing the
Left
and
Right
buttons
of the DADS Terminal.
Refer to End Condition of a Measurement
Recording to get to know when the test is ended
by the application.
8.
Confirm
Name of the
test or "No
tests"
Confirm by pressing the
Enter
key of the DADS
Terminal, or by clicking
Cancel
button on the main
screen.
Remarks:
Qualities
have to be defined before test can be created with the test list editor.
When the torque becomes higher than a certain value (
Torque Limit
in HITEC
Config), during a measurement, the DADS terminal stops automatically the motor.
This is done for security reasons, independently of the application running on the
PC. The application will not go further until the torque becomes lower than this limit.
In this case the application seems to be blocked. In a normal situation this doesn't
happen, as the torque becomes low immediately after the motor stopped.
TLS and Normalization Test
These measurements can be done by choosing the menu Maintenance / TLS or
Maintenance / Normalization. The measurement is done exactly like a normal one, with the
following difference. The name of the test and the test itself has not to be defined by the
user. This is done by the application.
After having selected the grade to make a new test, the user has to activate the test and
start by pressing the motor key of the Absorptometer.
Abort a Test
During the measurement a test can be stopped by stopping the motor (pushing the Stop key
on the DABS terminal). In this case the terminal checks the validity of the measurement. If
the measurement is valid, the operator will be asked to save the measurement. "
Save?
" will
be displayed on the terminal LCD. Press the
Enter
key on the DADS Terminal to save the
test, or the
Left
key to abort the test.
In case the
Enter
key has been pressed, "
Save
" will appear on the terminal LCD and the
operator has to select the part of the test he wants to save, by using cursors on the
application's main graph. If the selection is made the
Save button
has to be clicked, and the
graph will be saved. In case the
Left
key is pressed, the test is deleted.