Torque
The value of one torque unit is calibrated for type E and type H
Absorptometer in mNm (1/1000 Nm) - for the type B model it is
not clearly defined because the adjustment does not respect a
physically absolute procedure (see appropriate instructions).
The available torque measuring range is 0 to 100 for type B
and type E models and 0 to 150 for type H model. However,
on the type H, the torque measuring range can be limited to
100 units in order to make it behave as type B or E.
Run-OUT Time
Time in seconds to shut-off the run, after a maximum has
been observed in the raw data curve.
Threshold [%]
After a stop condition occurred, the system goes back to the
recorded raw data and determines the moment when this
value, in % of the highest observed raw data, occurred and it
takes this moment as the beginning point of the data set to
calculate the polynomial.