Proceed like follows:
1)
Push the
Check
button to start the test.
2)
"
Rise
" should be displayed in the terminal window now.
Apply a force
to the torque
sensing system until the message "
Release
" will be displayed on the terminal screen.
3) Remove the force
. The actual torque value will be displayed in the main graph, like in
the example below.
In case the machine is adjusted in the right way the attenuation time should fall in the
marked region around 3 s (3 ± 0.2).
In case the attenuation doesn't fall within the specifications the machine should be re-
adjusted and the test should be repeated from point two on.