The
button allows printing out the graph results, including graph and configuration
settings. This button has the same function then the menu File / Print Setup.
Configuration Section
The Configuration section provides some information about the actual configuration of the
test system terminal during the test. In case the application is in the retrieve modus the latest
configuration of the test is shown. This
is due to the fact that the settings can
change during a test, by using a non
standard sequence. The following
information is provided.
TLS
The actual TLS during the test. See also Definitions and Terms for more
information about the TLS.
Chamber
The chamber with which the test is made.
Burette
The speed of the burette (ml/min).
Liquid
Displays the liquid used for the absorption test.
Motor speed
The motor speed (rpm).
Sampling rate
The sampling rate of the terminal (samples/10 ms).
Graph Section
This part of the main
window displays the
torque
vs. the quantity
graph, of
the test made, or the actual
value of, in case the test is
running.
The
scale
of the torque can
be changed, by pressing on
the
arrow buttons.
The blue curve indicates
the raw torque curve. The
green one shows the
calculated polynomial
curve. And the yellow line
indicated the current TLS
value.
By "
double-clicking
" on
the graph window the graph
X scale sets itself to make
the complete graph visible.