Backup Config
This part of the application gives the user the possibility to make
backups
of the current
configuration. Backups should be done in a regular manner. Like this no major problems will
appear in case there are problems with the PC hardware (broken disk ...). See also the FAQ
/ Frequently Asked Questions - How to copy the DADS application configuration from one
PC to another one.
Backups are done by using the HITEC config application. If the user selects the backup
menu the HITEC Config application is started in the backup operator mode. The user has the
possibility then to generate backups.
In case the user wants to do a restore, he has to log on into the HITEC Config application as
Administrator. In this case a password is required.
Refer to the HITEC Config user manual,
, for more information on how to do a
backup and restore.
Pre-polish
This operation is available only for machine H type Absorptometers.
New chambers have to be pre polished, according to the ASTM specifications D2414,
annexe 3, before they can be used for absorption measurements.
Proceed like follows for the pre polish of a chamber.
1. The following window appeared on the
screen after the user chooses the pre
polish menu.
Follow the instruction and fill in 25 g of
carbon black SRB-F. Start the process by
pressing the
Motor Start
button on the
bottom side of the type H Absorptometer.
2. The terminal will fill 35 ml of liquid and then continue mixing during 16 hours. The
following window is displayed, showing the
elapsed time.
3. The user can open the door or stop the
motor at any time during the pre-polish and
continue the pre-polish by restarting the
motor pressing the
Start
button on the
equipment.
4. During a pre-
polish, the terminal displays “PP 17/May 15:36 xxxx” where “17/May
15:36
” is the date when the pre-polish will be finished and “xxxx” is the torque value.
If the user opens the door or stops the motor this date/time is automatically
incremented until the motor starts again.
The user can cancel the pre-polish on the terminal by using the
Left
button and then the
Enter
button to validate.
Remark:
The torque is displayed on the right side of the display. It will go down during the
pre polishing process.