FAQ / Frequently Asked Questions
Fo
llowing a list with frequently asked questions and, and their answers, is shown. Some
common problems are given too.
1.
Question
How to copy the DADS application configuration from one PC to
another one
.
Answer
Backup the current configuration to a drive on the network (or floppy disk);
by using the Backup function in the HITEC Config application.
Install the basic application on the second PC.
Restore the configuration from drive on the network (or floppy disk), by
using the Backup function.
2
.
Question
After start of the measurement the torque is shown, but the liquid
quantity doesn't change.
Answer
Check the burette type. It might be that you are running with a DADS
Version 1.0 terminal and using the pulse mode burette. The burette has to
be one of the other types.
3
.
Question
The motor is running but no liquid is coming out of the burette. No
control over the burette in the Maintenance/burette purge menu.
Answer
Check the burette type and the burette cable.