Hitec Luxembourg S.A. |
5, Rue de l’Église, L-1458 Luxembourg
www.hitec.lu | [email protected] | Tel.: (+352) 49 84 78 - 1 | Fax. (+352) 401303
File Name: HTCA_003-DADS-UserManual.docx
Document Type: End User Manual | Version date: 01/05/2010 | Last Printed: 15/02/2011
Page 50/81
Burette Debit
The burette Debit test has two
main goals
:
For DADS terminals using
a
constant rate burette
.
Determine the real debit of the burette. This value will be stored
and all result calculations depend on this value.
For DADS terminals using
a
pulse mode burette
.
Check whether the control is all right. The theoretical filled liquid
(the one the application wants to fill) at the end of the test
should correspond to the value indicated on the burette.
Proceed like follows to make the
verification
.
1.
Weight the recipient
the liquid will be filled in. Set the balance to zero, while
weighting the recipient.
2.
Enter the time
you want the burette to fill. 360 seconds is default but should be
considered as a minimum. Push on the OK button to start the burette filling out.
3. The
burette will fill
at a constant rate of
4 ml/min
during the selected time.
4. At the end of the filling time you have to enter the following values in the associated
textbox:
Weight of the recipient with the liquid
and insert the weight;
Enter the
temperature of the liquid
;
Enter the
liquid quantity
given on the display of the burette.
5. Click on the
OK
button after the values are entered.