3 - 43
(c) The recorder speed setting screen appears.
<Recorder speed setting screen>
RCD SPEED (40 nm/min=1, 60 nm/min=2)
2
Press numeric key 1 or 2 and then the ENT key to set
a recorder speed.
* Initial (default) value: 2
(d) When the setting of recorder output speed is finished,
the system returns to the 1
st
monitor screen.
(3) Setting of Output Spectral Data
(a) Press the RECORD
key.
(b) The recorder setting screen appears.
<Recorder setting screen>
RECORD (RCD RANGE=1, RCD SPEED=2,
3
SPECT RCD=3)
Press numeric key 3 and then the ENT key to select
SPECT RCD (spectral recording).
(c) The output data spectrum No. setting screen appears.
<Output data spectrum No. setting screen>
RECORD SPECTRUM NO. (1 - 4)
1
Press a numeric key 1 to 4 and then the ENT key to
set a spectrum No. to be output.
* Initial (default) value: 1
(d) The background spectrum No. setting screen now
appears.
The background spectrum is used to compensate for
absorption by the solvent or cell. To ensure that an
appropriately corrected spectrum is used in analysis, a
new background spectrum should be acquired
whenever the solvent or wavelength scan range is
changed.