APPENDIX 1 - 14
6.4 System Tubing
In the Primaide1000 series, the inlet tube up to the pump and
drain tube are fastened to the right/left side face of each unit with
tube clamps. And the tubes are connected from the pump to
the autosampler, column and detector through the central hole of
each module unit.
(1) Isocratic System (stacked in one block)
Connect the inlet tube of mobile phase to the IN port of the
degassing device. Connect the OUT port of the degassing
device to the inlet of the pump. Connect the drain valve of
the pump to port 1 of the injection valve of the autosampler.
Connect port 2 of the injection valve to the column inlet.
Connect the column outlet to the inlet tube of the detector
cell. For washing solution, provide connection to port 1 of
the syringe valve of the autosampler through the degassing
device.
Drain tray
1
1
6
110
OOT
3
2
Inlet tube
Mobile Pashe
Column
Injection valve
Drain valve
Detector cell
Drain
Primaide1210
Autosampler
Primaide1310
Column Oven
Primaide1410/1430
UV/DAD Detector
Primaide
Organizer
Primaide1110
Pump
Washing Solution
Syringe valve
Degassing
device
Drain
OUT
IN