5 - 1
5. SPARE PARTS
5.1 Replacement Parts and Consumables
Table 5-1 lists the replacement parts and consumables for this
instrument. It is suggested that a suitable quantity of each item
be stocked in the laboratory to minimize down time. For many
of these items (e.g. lamps), the quantity is dependent on the
level of use of the instrument.
The estimated service life of each part is based on the
assumption that the instrument is operated 160 hours/month
(8 hours/day), 1920 hours/year (240 days).
Table 5-1 Replacement Parts and Consumables
Part Number
Part Name
Estimated Service Life
Remarks
650-1500
150 W Xenon
lamp
Guaranteed life : 150 hours
Limit
: 500 hours
Warranty period : 6 months after
delivery to
customer
890-2019
Long-life xenon
lamp (Special
accessory)
Guaranteed life : 500 hours
Limit
: 1000 hours
Warranty period : 6 months after
delivery to
customer
890-2662
Mercury lamp
Guaranteed life: 100 hours/
2 years
J821399
Time-lag fuse
5 A
890-2932
Flow cell (12
L)
9600 hours/5 years
890-2684
Drain tube
ID 0.33
0.62 T
9600 hours/5 years
Length: 2 m
Outside diameter:
1.57 mm
893-5860
Inlet tube
ID 0.25
0.66 T
9600 hours/5 years
Length: 60 mm
Outside diameter:
1.57 mm (inlet
ferrule included)
L369053
Ferrule front
9600 hours/5 years
Part for connecting column
and inlet tube
Material: Daiflon
M790019
Screw S (SUS)
9600 hours/5 years
Example use:
Connecting pipe
893-0827
Connecting pipe
(SUS, coated)
9600 hours/5 years
Example use:
Piping between
column outlet and
inlet tube
For standard flow cell
Length
: 350 mm,
Inner diameter : 0.25 mm,
Outer diameter : 1.57 mm,
Material
: SUS,
PTFE (for
coating)
Содержание PRIMAIDE 1440