4 - 6
(x) The 1
st
monitor screen returns.
(d) Setting of spectrum measurement:
(i)
Press the SPECT key. The scan mode setting
screen appears.
<Scan mode setting screen>
SPECTRUM MODE (FL (Ex)=1, FL (Em)=2)
2
Press numeric key 2 and then the ENT key.
(ii) Set a scanning range.
➀
Enter a starting wavelength (minimum value)
for WL1.
<Scan range setting screen>
SCANNING WL RANGE (WL1 WL2) nm FL (Em)
<250
– 850>
350 - 700
Press numeric keys 3 5 0 and the ENT key.
➁
Enter an end wavelength (maximum value) for
WL2.
<Scan range setting screen>
SCANNING WL RANGE (WL1 WL2) nm FL (Em)
<
400 - 750
>
350 - 450
Press numeric keys 4 5 0 and the ENT key.
(iii) The memorize spectrum No. setting screen
appears.
<Memorize spectrum No. setting screen>
MEMORIZE SPECTRUM NO. (1 - 4)
1
Press a numeric key (1 to 4) and the ENT key.
(iv) The scan monitor screen appears, and spectrum
Содержание PRIMAIDE 1440
Страница 1: ...PRIMAIDE 1440 荧光检出器 ...