3 - 6
3.1.3 Auto Zero adjustment
The output value should be zeroed when zero point adjustment
is required or when the data baseline deviates from the recorder
zero point.
(1) Press the
A/Z
key.
(2) The FL indication on the 1
st
monitor screen is reset to zero
and output signal becomes zero.
<The 1
st
monitor screen>
TIME
Ex
Em
FL
PROG
LAMP
[I]
0.0
250
350
0.0
1
Xe
* Setting value range: 0 to 1000
If the range is exceeded, an error message is displayed
3.1.4 Wavelength setting
Specify Ex (Excitation wavelength) and Em (Emission wavelength).
(1) Press the
WL
key.
(2) The following display will be presented. Set an excitation
wavelength (Ex) (cursor appears under Ex).
<Ex wavelength setting screen>
Ex
Em
(200
– 850,0) nm
250
350
Press numeric keys (2 0 0 to 8 5 0) and then the ENT key.
* This wavelength is settable within a range from 200 to
850 nm in increments of 1 nm.
* Default value: 250 nm
* Press the ESC key or
key to cancel the input excitation
wavelength value. The 1
st
monitor screen returns.
* Pressing the
key enables you to advance to the
emission wavelength setting without changing the
excitation wavelength.