3 - 25
* If ESC key is pressed prior to ENT key, or if this status
continues for 5 minutes or longer, the system will return
automatically to the 1
st
monitor screen.
(5) The result of emission wavelength (Em) check will now be
displayed.
<Em side check result>
Em SIDE ACCURACY (254 nm)
GOOD -1nm
Press the ENT or ESC key.
* If the error is within
3 nm, “GOOD” will appear, and if the
error exceeds
3 nm, “FAIL” will appear on the display.
* If 5 minutes or more elapses without pressing the ENT or
ESC key, the system will return automatically to the 1st
monitor screen.
(6) When the wavelength accuracy check is finished, the
system returns to the 1
st
monitor screen.
3.3.7 Xe lamp energy check
Energy level of the xenon lamp can be checked using the xenon
lamp beam at 254 nm.
(1) Press the
GLP
key.
(2) The following display will be presented.
<GLP initial screen>
GLP
(LOGBOOK=1, KEY LOCK=2,
4
WL CHECK=3, Xe CHECK=4)
Select Xe CHECK.
Press numeric key 4 and then the ENT key.
* If the Xe lamp is turned off, the system will return to the
1
st
monitor screen without performing an energy check.
(3) The Xe lamp energy check screen appears and an energy
check will be performed.
Содержание PRIMAIDE 1440
Страница 1: ...PRIMAIDE 1440 荧光检出器 ...