Chapter 9 PLC Operation
9-5
9.1.2
Periodical Scan
(1)
Definition and operation
This scan executes interrupt programs (periodical scan programs) while the CPU is operating with a fixed cycle time (10
ms, 20 ms, or 40 ms) specified by the users.
Enter the periodical scan program to be executed between instructions INT0 and RT1 if it should be started up with a 10
ms cycle time, and between INT1 and RT1 if it should be started up with a 20 ms cycle time.
The periodical system processing is executed every 10 ms regardless of whether or not there is a periodical scan program.
Figure 9.9 Operation of periodical scan (in case of INT1)
(2)
Causes of congestion errors at periodical scan
If there are periodical scans at every 10 ms as well as scans at every 20 ms or 40 ms, a congestion error occurs and the
scan is stopped if the periodical scan at 10 ms is started up again before all the periodical scans are completed (i.e., the
periodical system processing at INT0 to INT2 does not end within 10 ms).
Figure 9.10 Congestion error at periodical scan (10 ms)
Similarly, when executing with a periodical scan at every 20 ms or with a combination of periodical scans at every 20 ms
and 40 ms, a congestion error occurs if the periodical scan at 20 ms is started up again before all the periodical scans are
completed (i.e., the periodical system processing at INT1 to INT2 does not end within 20 ms). Finally, when using a
periodical scan at every 40 ms, a congestion error occurs if the periodical scan at 40 ms is started up again before all the
periodical scans are completed (i.e., the periodical system processing at INT2 does not end within 40 ms).
RTI
INT1
10 ms
20 ms
RTI
INT1
INT1
Normal scan
Scan every 40 ms
Periodical program
Scan every 20 ms
Periodical program
Scan every 10 ms
Periodical program
Periodical system
processing
Periodical scan
Program execution
Periodical scan
Normal scan
Periodical system processing
(including interrupt analysis
processing)
Interrupt
Interrupt
Interrupt
INT0
10 ms
10 ms
INT1
INT0
RTI
10 ms
INT2
INT1
INT0
RTI
RTI
END scan processing
Normal scan program
Stop
Periodical scan
Normal
scan
(Example 3)
Before the periodical scan program
at every 40 ms ends, the periodical
interrupt at 10 ms is started up again.
(Example 2)
Before the periodical scan program at
every 20 ms ends, the periodical
interrupt at 10 ms is started up again.
(Example 1)
Before the periodical scan program at
every 10 ms ends, the periodical
interrupt at 10 ms is started up again.
Scan every 10 ms
Periodical program
Scan every 20 ms
Periodical program
Scan every 40 ms
Periodical program
Periodical system processing
Stop
Stop
Содержание HIDIC MICRO-EH
Страница 1: ...HITACHI PROGRAMMABLE CONTROLLER APPLICATION MANUAL NJI 350B X ...
Страница 12: ...MEMO ...
Страница 14: ...Chapter 1 Features 1 2 MEMO ...
Страница 50: ...Chapter 4 Product lineup and wiring 4 18 MEMO ...
Страница 196: ...Chapter 5 Instruction Specifications 5 146 ...
Страница 263: ...Chapter 11 Communication Specifications 11 10 MEMO ...