![Hioki MR8880-20 Скачать руководство пользователя страница 281](http://html1.mh-extra.com/html/hioki/mr8880-20/mr8880-20_instruction-manual_2137332281.webp)
Appendix 7 Supplemental Technical Information
A
19
Appendix
The input signal acquisition rate corresponds to time
per division on the horizontal axis.
Determining the timebase: Calculate from the fre-
quency and period.
f [Hz] = 1/t [s] (f: frequency, t: period)
Example: If the measurement frequency is 50 Hz
50 [Hz] = 1/t [s] t = 1/50 [s] = 0.02 [s] = 20 [ms]
To enable 1 period to be displayed with 10 divisions,
the timebase becomes 20 [ms]/10 [div]= 2 ms/div.
To display multiple periods, set a timebase range that
is slower than 2 ms/div.
From the selection ranges for the timebase range, se-
lect a range that is close to the calculated value.
To measure phenomena with relatively fast signals
such as instantaneous waveforms, we suggest setting
a small value (if the frequency is 50 Hz, the timebase
should be set faster than 5 ms/div).
During and after measurement, waveforms can be ex-
panded and compressed along the time axis.
This instrument converts analog input signals into
digital values which are then processed internally as
digital (numerical) values. This A/D conversion pro-
cess is called sampling. Sampling repeatedly mea-
sures the size of the input signal at a specific interval
(the sampling period).
The rate of measurement is called the sampling rate.
Sampling units are [S/s] (read as samples-per-sec-
ond)
This is the number of samples taken each second,
and is the inverse of the sampling period. (1/T)
Appendix 7 Supplemental Technical Information
Selecting a Timebase (Horizontal Axis)
20 div
Timebase: with 1 div = 5 ms
Timebase: with 1 div = 10 ms
1 div
Timebase and Sampling
Sampling Period
Sampling Points
Measurement
Содержание MR8880-20
Страница 2: ......
Страница 12: ...Procedure 2 ...
Страница 40: ...1 3 Screen Configuration and Display 30 ...
Страница 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Страница 174: ...6 4 Data Management 164 ...
Страница 186: ...7 7 Performing a Printer Check 176 ...
Страница 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Страница 216: ...10 4 Making System Settings 206 ...
Страница 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Страница 226: ...11 3 Communication Using Commands 216 ...
Страница 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Страница 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......