![Hioki MR8880-20 Скачать руководство пользователя страница 247](http://html1.mh-extra.com/html/hioki/mr8880-20/mr8880-20_instruction-manual_2137332247.webp)
13.3 Real-time Function
237
Cha
p
te
r 1
3 S
p
ecifi
cations
13
Screen settings
Single, Dual, Quad
Waveform compres-
sion and magnifica-
tion
Horizontal axis (time axis):
10ms, 20ms, 50ms, 100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 30s/
div, 1min, 2min, 5min, 10min, 30min, 1h, 2h, 5h, 10h, 12h, 1day/div
Vertical axis (voltage axis): Differs depending on display method
setting.
• When set by position:
Magnification: x20, x10, x5, x2, x1, x1/2, x1/5, x1/10
Position: Set -50 to 150% (when magnification x1)
• When set by upper and lower limits: Set the upper and lower limits.
The display is as follows when the measurement value is over the
set range.
• "OVER" when over on + side
• "-OVER" when over on - side
Display screen
Waveform display
Simultaneous waveform and trigger setting display
Simultaneous waveform and setting display
Simultaneous waveforms and numerical calculation results display
Waveform and A/B cursor values displayed at the same time
Waveform scrolling
Scrolling possible in time axis direction only for data portion stored
in internal buffer memory
(Possible both during measurement and after measurement)
<<
/
>>
These keys make large jumps
Jump function
Specifying and jumping to waveform display position possible.
(Specify a jump position in the waveform position display bar, and
then jump to the specified location.)
Real-time saving
Selectable from off, Waveform (realtime), CSV (realtime), Calc
(after meas.), Wa Calc, CSV + Calc
Saving simultaneously with measuring, drawing, and printing is
possible (limited by time axis).
Split save
Selectable Off, On or Periodic
• On: Saves data split into separate files from measurement start,
at the set interval.
• Periodic: Saves data split into separate files from a set reference
time (within 24 hours), at the set interval.
*Even if split saving is set to Off, the file will automatically be split if its
size is expected to exceed 500 MB.
Overwriting save
Selectable Off or On
• On: When there is not much space left on the storage media, the
oldest waveform file with the same extension is deleted and a
new file is created.
• Off: Data is saved until the storage media is full and then mea-
surement ends.
Содержание MR8880-20
Страница 2: ......
Страница 12: ...Procedure 2 ...
Страница 40: ...1 3 Screen Configuration and Display 30 ...
Страница 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Страница 174: ...6 4 Data Management 164 ...
Страница 186: ...7 7 Performing a Printer Check 176 ...
Страница 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Страница 216: ...10 4 Making System Settings 206 ...
Страница 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Страница 226: ...11 3 Communication Using Commands 216 ...
Страница 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Страница 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......