![Hioki MR8880-20 Скачать руководство пользователя страница 252](http://html1.mh-extra.com/html/hioki/mr8880-20/mr8880-20_instruction-manual_2137332252.webp)
13.4 Other Functions
242
External trigger ter-
minal functions
The external trigger terminal function is selectable from trigger and
event.
External trigger filter
Selectable from On and Off.
External input termi-
nal functions
Any of the following functions can be set for each of terminal 1 and
terminal 2.
Start, stop, abort, save, print
External output ter-
minal functions
Any of the following functions can be set for the terminal.
Error, BUSY, Waiting (waiting for trigger)
Open-drain output (with 5 V output, Active Low)
Output voltage level HIGH: 4.0 to 5.0 V, LOW: 0 to 0.5 V
Time value display
Select from time, date, and number of data for the horizontal axis
display.
The time, date, or number of data from the start of measurement is
displayed.
• Time: Displays the time (in hours, minutes, and seconds) from
the start of measurement.
However, if a trigger is set, the time from the trigger point is dis-
played.
• Date: Displays the date and time that the data was acquired.
• Number of data: Display the number of data from the start of
measurement.
However, if a trigger is set, the number of data from the trigger
point is displayed.
This setting is reflected in the values saved in CSV format.
Key-Lock function
Disables key operations
(Hold cursor left and right keys for three second to enable/dis-
able.)
System reset
All setting items can be restored to their factory defaults.
Beeper
On/Off
Auto range
Available (The most suitable time axis and voltage axis for the
input waveform are selected automatically.)
Screen hard copy
The screen image can be printed by pressing the COPY key on
the MR9000 Printer Unit.
List print
Outputs the setting conditions after the waveform data is printed.
(When List or List & Gauge is selected for the print setting condi-
tion)
In addition to from the display screen, output can be performed by
pressing the PRINT key on the MR9000 Printer Unit.
Printing density
Print density selectable from the following.
Dark, standard, and fine
Grid types
Type of grid to draw on the recording paper selectable from the fol-
lowing.
Off, normal, fine, normal (dark) or fine (dark)
Self-Test function
KEY/LED, LCD, ROM/RAM, and USB checks are possible.
Содержание MR8880-20
Страница 2: ......
Страница 12: ...Procedure 2 ...
Страница 40: ...1 3 Screen Configuration and Display 30 ...
Страница 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Страница 174: ...6 4 Data Management 164 ...
Страница 186: ...7 7 Performing a Printer Check 176 ...
Страница 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Страница 216: ...10 4 Making System Settings 206 ...
Страница 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Страница 226: ...11 3 Communication Using Commands 216 ...
Страница 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Страница 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......