Appendix 1 Error Messages and Remedial Actions
A
2
When there is a warning, a warning message will appear just once and then disap-
pear after a few seconds. A displayed warning message can be closed by pressing
any key.
015
The possible zero-adjustment range
has been exceeded.
INSTNT error: ....
RMS error: ....
There is a problem with the input section. If the er-
ror message appears after turning the power off
and then back on, the instrument needs to be re-
paired.
030
The printer has been stopped be-
cause a head temperature error was
detected.
Wait until the temperature of the printer falls be-
fore using the printer
031
The printer has been stopped be-
cause a motor temperature error was
detected.
Wait until the temperature of the printer falls be-
fore using the printer
032
The printer was stopped for an error
of the motor.
There is a problem with the printer motor. If the er-
ror message appears after turning the power off
and then back on, the printer needs to be re-
paired.
Message
Remedial Action
Warning Messages
Message
Remedial Action
101
Invalid key.
Keys are disabled while measuring. Try operating
again when measurement is finished.
102
Cannot start measurement.
The
START
and
STOP
keys are disabled when the
current screen is displayed.
103
Cannot change while measuring.
Press
STOP
to stop measuring before changing the
setting.
104
Voltage drop and wave compari-
son triggers cannot be modified.
(The effective time axis range is
100
μ
s/div to 20ms/div and the ef-
fective recording interval is 1
μ
s to
200
μ
s.)
Set the time base range or recording interval to a val-
id range.
105
Voltage drop and wave compari-
son triggers have already been
set and cannot be modified.
(The effective time axis range is
100
μ
s/div to 20ms/div and the ef-
fective recording interval is 1
μ
s to
200
μ
s.)
Disable the voltage drop and waveform judgment
triggers.
106
This cannot be used in RMS mode
(voltage drop and wave compari-
son trigger).
Change the measurement mode of the correspond-
ing channels to instantaneous value.
201
Exceeding the setting range.
The entered numerical value exceeds the setting
range. Enter an appropriate value.
Содержание MR8880-20
Страница 2: ......
Страница 12: ...Procedure 2 ...
Страница 40: ...1 3 Screen Configuration and Display 30 ...
Страница 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Страница 174: ...6 4 Data Management 164 ...
Страница 186: ...7 7 Performing a Printer Check 176 ...
Страница 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Страница 216: ...10 4 Making System Settings 206 ...
Страница 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Страница 226: ...11 3 Communication Using Commands 216 ...
Страница 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Страница 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......