11.4 Specifications of CAN Setting Function (for Memory HiCorder)
General specifications
Functions
Display function
1. CAN definition data list display (of all CAN definition data setting items)
2. 8910 registration list display (item no., input port no., signal name, label
name, unit, data length)
3. Analog output list display (ON/OFF, channel no., signal name, label
name, unit, D/A conversion format)
4. Logic output list display (ON/OFF, channel no., signal name, label name,
bit position, ID trigger setting)
Storage
1. CAN definition data (binary format, with extension "cdf")
2. 8910 setting data (all settings except for CAN definition data) (binary
format, with extension "cmu")
Communication function
Transfer of setting data to the 8910 via RS-232C (8910 setting data cannot
be read.)
Transfer of setting data to Memory HiCorder via media (Memory HiCorder
setting data cannot be stored on media.)
Model compatible with
communication control
8910 CAN ADAPTER, 1 instrument (effective only when the 8910 is in
ONLINE mode)
11.4 Specifications of CAN Setting Function
(for Memory HiCorder)
Type
CD-R, one disc (including "8910 CAN set program")
Supported measuring equipment 8826, 8841, 8842 (9557 RS-232C CARD required)
Settings
1. Output channel setting
(1) Allocation of 8910 registration list to analog output channels (12
channels)
(2) Allocation of 8910 registration list to logic output channels (24 bits)
2. D/A conversion format (analog channels only)
(1)Selection of bipolar (output of -5 V to 5V) or monopolar (output of 0 V
to 5 V)
3. Bit position setting (logic channels only)
4. Memory HiCorder channel setting (analog channels, logic channels)
(1)Correlation between 8910 output channels and Memory HiCorder
input channels
Display function
8910 setting data display
Storage
8910 setting data x 6 blocks can be stored in internal backup memory of the
Memory HiCorder.
Simultaneous storage upon Memory HiCorder channel setting (Data
cannot be stored on media.)
Communication function
Transfer of setting data to 8910 via RS-232C (8910 setting data can also
be read.)
Transfer of setting data from PC via media (Memory HiCorder setting data
cannot be stored on media.)
Automatic scaling function
Automatic calculation of Memory HiCorder scaling setting based on 8910
setting data
Содержание 8910
Страница 1: ...8910 Instruction Manual CAN ADAPTER EN Aug 2018 Revised edition 7 8910A981 07 18 08H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 18: ...Notes on Use 10 ...
Страница 44: ...3 5 Operation Map 36 ...
Страница 48: ...4 2 Program Setup 40 ...
Страница 56: ...5 4 Operation Flowchart 48 ...
Страница 78: ...6 6 Editing in Another Window 70 ...
Страница 90: ...7 4 Setting the ID Filter 82 ...
Страница 99: ...91 9 2 Setting Data Printout ...
Страница 130: ...10 2 Using the 8910 with the 8841 8842 MEMORY HiCORDER 122 ...
Страница 147: ......
Страница 148: ......
Страница 149: ......
Страница 150: ......