11.1 Messages
______________________________________________________________
93 Disk full.
Little capacity is available.
Delete files or use new media.
94 Path name error.
Up to 127 characters can be used for
a path name.
95 Empty directory name.
Name the directory.
96 Directory already exists.
Another directory of the same name
exists. Saving cannot be performed.
99
Conditions for OVERWRITE
are not satisfied.
Match the unit condition and file data
condition (function and time-axis).
201 Set printer paper.
Printer paper has run out. Reload.
202 Set printer lever.
The head up/down lever has been
left in the up position.
205 Invalid.
The key pressed is not valid,
because measurement operation is
in progress.
300 Cannot START.
Cannot start measurement from
Advanced setting screen
349 Invalid.
When the time axis range is 20
ms/DIV, the waveform evaluation
trigger is disabled.
382 No waveform data.
There is no waveform data to display
on the Waveform Display screen.
Data must be input before
processing can start.
397 Out of range.
Input the appropriate value.
398
A/B cursor positions
invalid.
Move the cursors to the appropriate
position.
620 Battery low.
Recharge or change the battery
pack.
621 Battery low
Battery voltage is low.