3.5 Trigger Functions
______________________________________________________________
Rising trigger slope :
Filter width (0.5 DIV)
Trigger level
Triggering does not occur here
Trigger point
0 V
-600 mV
B
0 V
A
Point A
Trigger level: 200 mV,
Trigger slope: rising ( )
Point B
Trigger level: -600 mV,
Trigger slope: falling ( )
200 mV
Trigger Filter
Triggering occurs when the trigger conditions are met
within the filter width (0.5 DIV). This is useful to prevent
spurious triggering by noise
To cause triggering at point A or point B with the sine
wave shown below, make the following settings.