![Hioki 8715-01 Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/hioki/8715-01/8715-01_instruction-manual_2137217015.webp)
vii
_____________________________________________________________________
Safety Notes
______________________________________________________________
f.s.
(maximum display value or scale length)
The maximum displayable value or the full length of the
scale.
This is usually the maximum value of the currently
selected range.
rdg.
(reading or displayed value)
The value currently being measured and indicated on the
measuring instrument.
dgt.
(resolution)
The smallest displayable unit on a digital measuring
instrument, i.e., the input value that causes the digital
display to show a "1".
Accuracy
We define measurement tolerances in terms of f.s. (full
scale), rdg. (reading) and dgt. (digit) values, with the
following meanings: