3.7 Start and Stop Operation (Waveform)
______________________________________________________________
Pressing the START key causes a
confirmation window to appear.
Press the START key again to start
measurement.
The LED lights green when measurement
starts.
When pre-trigger is activated, the trigger is
not enabled until the specified pre-trigger data
has been read.("Pre-trig wait" is displayed.)
"Wait for trig" is displayed until the trigger
condition occurs.
Data recording starts when the trigger
conditions are met.
If no trigger conditions are enabled, data
acquisition starts when measurement starts.
During recording, "Storing" is displayed.
When automatic printing is selected, data is
printed after display.
Measurement can be aborted by pressing the
STOP key twice.
Waveform data acquired before the Stop key
was pressed is displayed and saved.(automatic
printing will not execute)
測定開始
Measurement start
Trigger conditions
are met
測定終了
Measurement end
Recording data
Wait for trigger
3.7 Start and Stop Operation (Waveform)
・
If measurement is aborted before a trigger event has
occurred, no waveforms are stored. However, if a
waveform was acquired during a previous trigger event,
that waveform is displayed (unless it was the maximum
recording length).
・
When the number of recorded trigger events has reached
the sequence limit, the earliest recorded data is
overwritten.