5.4 DMM Function
______________________________________________________________
Procedure
Screen: Waveform Display
Display hold
Press the STOP key to Hold the
current status and "HOLD" is
displayed.
Press the START key to cancel
the hold.
DMM Screen Printing
・
Pressing the PRINT key
prints the currently
displayed values.
・
Pressing the COPY key
copies the screen.
Measuring object
Commercial mains (50/60 Hz, auto-selection)
Displays
RMS value
Update rate
1 s (Sampling rate: 4 kS/s)
Number of displayed digits 4 (The lowest digit is displayed as 0 for values
0 to 4, and as 5 for values 5 to 9.)
Accuracy
3% rdg. 5 dgt.
NOTE
The display changes to the next lower range when the count is
less than 500. The minimum digit display is either 0 or 5.
The displayed color of the digital values for each channel
corresponds to the waveform display color.
5.4 DMM Function
Commercial mains can be displayed RMS values in real
time. Correct values are displayed only when the input
voltage is commercial mains (50/60 Hz).
1. Press the DMM key to display the DMM screen.
Measurement begins when the DMM screen appears.
2. When finished with the DMM screen, press any screen
display key other than DMM.