8.3 Settings on the File Screen
______________________________________________________________
Load
Loaded part
Load
NOTE
Loading only a Trend RMS Recording file (RMS extension)
erases the instantaneous value data in internal memory. To load
both instantaneous and rms values at the same time, load the
index file (RMM extension), or load the rms file first. Also, the
loading capacity for instantaneous value data depends on the
settings of the instrument at load time.
If the recorded length of the saved data is longer than the
recording length set in the 8715-01, the saved data is
loaded for the set recording length, and the remainder of
the data is not loaded.
If the recording length set in the 8715-01 is longer than the
saved data, the data from the end of the last saved
waveform to the end of the 8715-01 recording length is
invalid.
(5) Press the START key to load, and press the STOP key to
cancel.