
Chapter 4: Software Interface
36
Exicor GEN5 User Manual
Clicking on the
Offsets
button performs a single point measurement at the current x-
y position to use as a baseline for sample measurements. There should be no
sample in the beam path when pushing the
Offsets
button. Clicking on the
Home &
Reset
button reinitializes the system hardware. The
Notes
button is used to add
comments to the current data file. The
Statistics
button shows the birefringence
magnitude and fast axis angle statistics for the current scan.
Figure 4.11 Offsets, Home & Reset, Notes
Figure 4.12 shows the status bar located at the bottom of the Exicor GEN5 main
window.
Figure 4.12 Sample Thickness Indicator
This bar shows the sample thickness that has been entered by the user. It is
important to have an accurate sample thickness if you wish to plot the differential
stress in the sample. The sample thickness is for display only and can be changed
by starting a new scan and defining new sample thickness as described on page 24.
It can also be changed in the system settings window.
Содержание Exicor Gen5
Страница 1: ...USER MANUAL Hinds Instruments Inc P N 010 0000 061 UM Rev B ...
Страница 2: ......
Страница 3: ...USER MANUAL Hinds Instruments Inc P N 010 0000 061 UM Rev B Software 025 0000 066 Version 1 0 ...
Страница 10: ...Table of Tables viii Exicor GEN5 User Manual ...
Страница 22: ...Operators Safety Summary 12 Exicor GEN5 User Manual ...
Страница 72: ...Chapter 4 Software Interface 62 Exicor GEN5 User Manual ...
Страница 90: ...Chapter 7 Troubleshooting Help 80 Exicor GEN5 User Manual ...
Страница 94: ...Appendix A Technical Specifications 84 Exicor GEN5 User Manual ...
Страница 97: ......