2.65 Junction Board Debugging—IntelliDrive® Transport System
Chapter 2: Troubleshooting Procedures
connector pins at P6 on the Junction board. Also, check connectors P2 and P5
where the enable switches plug into the Junction board.
The combined enable switch signal connects to the PACM board via the
Junction-PACM board cable. To verify the signal is being received at the
PACM board, unplug the 14-pin cable from the PACM board and measure
continuity across pins 1 and 3. When either switch is depressed the switch
closure can be measured across the pins.
Throttle Debugging
The base part of the handle that connects to the frame contains a strain element
that provides an output signal proportional to the force applied to the handle.
The handles are very similar to the load beams used in the scale system. The
Junction board amplifies this signal and provides an output to the PACM
board.
Verify the output signal is correct at P6.10. P6. 4 is used as a ground reference
for measuring signals. When no force is applied to the handle the output signal
should measure 2.4V to 2.6V DC. The voltage output can be adjusted by
turning potentiometer R1 until the output signal measures 2.5V. When either
handle is pushed, the output signal will increase until it reaches 4.0V to 4.5V.
When either handle is pulled the output signal will decrease until it reaches
0.4V to 1.5V. This indicates the throttle circuit working correctly.
Before making any adjustments verify the supply voltages. The supply voltage
at P6.8 will measure more than 22 V on a working system. If no voltage is
present, or the battery voltage is low, go to the battery checkout procedure.
The excitation voltage, P3.5 and P7.5 (red wire) will measure 10V to 12V on a
working board. Also, the signal voltages at P3.3, P7.3, P3.4, and P7.4 will be
approximately ½ the voltage measured at P3.5 and P7.5. These are the green
and white wires on the handle connectors P3 and P7. If the excitation voltage is
lower than 10V, unplug the connectors P3 and P7 one at a time to see if the
voltage comes into range. If this occurs, one of the handles is probably
damaged and needs to be replaced. If the voltage never comes into range,
verify the 12V supply at U2.2 to P6.4 on the Junction board. If a bad strain
element is suspected, unplug the bad handle, readjust the zero output to 2.5V
with potentiometer R1. Verify the bed power unit operates using the good
handle. Replace the damaged handle.
2
VersaCare™ Bed Service Manual (MAN333 REV 2)
Page 2 - 79
Содержание VersaCare P3200
Страница 2: ......
Страница 6: ...Revisions NOTES Page iv VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 18: ...NOTES Page xvi VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 34: ...Warning and Caution Labels Chapter 1 Introduction NOTES Page 1 16 VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 270: ...Exchange Policy Chapter 5 Parts List NOTES Page 5 4 VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 272: ...Warranty Chapter 5 Parts List NOTES Page 5 6 VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 275: ...Recommended Spare Parts Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 9 ...
Страница 276: ...Base Frame Chapter 5 Parts List Base Frame Figure 5 2 Base Frame Page 5 10 VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 279: ...Base Frame Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 13 ...
Страница 281: ...Casters Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 15 ...
Страница 291: ...Head Siderail Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 25 ...
Страница 295: ...Siderail Center Arm Assembly Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 29 ...
Страница 299: ...Power Supply Sheet 1 of 2 Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 33 ...
Страница 306: ...Air Module Chapter 5 Parts List Air Module Figure 5 14 Air Module Page 5 40 VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 311: ...Sensor Strips Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 45 ...
Страница 312: ...Sleep Deck Chapter 5 Parts List Sleep Deck Figure 5 18 Sleep Deck Page 5 46 VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 325: ...Treatment Surface P3251EA Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 59 ...
Страница 337: ...Labels Sheet 2 of 2 Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 71 ...
Страница 338: ...CPR Cable Chapter 5 Parts List CPR Cable Figure 5 27 CPR Cable Page 5 72 VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 340: ...CPR Cable Chapter 5 Parts List NOTES Page 5 74 VersaCare Bed Service Manual MAN333 REV 2 ...
Страница 363: ......
Страница 365: ...Electronics Schematic P3200 Sheet 1 of 3 Chapter 3 ...
Страница 366: ...Electronics Schematic P3200 Sheet 2 of 3 Chapter 3 ...
Страница 367: ...Electronics Schematic P3200 Sheet 3 of 3 Chapter 3 ...
Страница 368: ...Motor Control P C Board Schematic PN 68717 Chapter 3 ...
Страница 369: ...Logic Control P C Board Schematic PN 69378 Chapter 3 ...
Страница 370: ...Siderail Controls Schematic Chapter 3 ...
Страница 371: ...Optional Controls Schematic Chapter 3 ...
Страница 372: ...Patient Controls Schematic Chapter 3 ...
Страница 373: ...Patient Pendant Schematic Chapter 3 ...
Страница 374: ...Power Supply PN 72289 Sheet 1 of 2 Chapter 3 ...
Страница 375: ...Power Supply PN 72289 Sheet 2 of 2 Chapter 3 ...
Страница 376: ...Motor Control P C Board PN 72273 Chapter 3 ...
Страница 377: ...Power Supply P C Board PN 70786 Sheet 1 of 2 Chapter 3 ...
Страница 378: ...Power Supply P C Board PN 70786 Sheet 2 of 2 Chapter 3 ...