Specifications
Chapter 1: Introduction
DuraStar™ Series Stretcher Service Manual (man268)
Page 1 - 13
1
Stretcher Back Section Inclination
Both models of the DuraStar™ Series Stretcher are mechanized so that the
attendant can raise or lower the back section by squeezing the release handle
located under the back section. The back section can be elevated from 0° to
90°.
Hilow
The hydraulic stretcher is mechanized so that the attendant can raise or lower
the sleep surface by pressing the appropriate pedals located on either side of
the stretcher. The stretcher is equipped with two hydraulic cylinders to ensure
that the stretcher lowers evenly with uneven loads.
Brake/Steer
Both stretcher models are mechanized so that the attendant can set the brake,
steer, or put the stretcher in the neutral position from anywhere around the
stretcher. The pedals are color-coded to enable quick use. All four casters
provide braking for the stretcher.
Trendelenburg/Reverse Trendelenburg
The hydraulic stretcher is mechanized to achieve the Trendelenburg/Reverse
Trendelenburg positions by pressing the proper foot pedals located on either
side of the stretcher. Quick Trendelenburg can be achieved by grasping the
foot end of the stretcher and pulling up. A maximum of 18°
Trendelenburg/Reverse Trendelenburg can be achieved with the stretcher
raised to the high position.
Siderails
Both stretcher models are equipped with 3/4 length siderails. They can be
moved from the stored position to the upright position in one motion.
The siderails should always be in a full upright position and latched when a
patient is unattended. When the siderails are raised, an audible click indicates
that they are completely raised and locked in place.
Siderails are intended to be a reminder, not a patient retaining device.
Appropriate medical personnel should determine the level of restraint
necessary to ensure a patient remains safely in bed. Consult the security strap
manufacturer’s instruction for use to verify the correct application of each
security device.
Содержание DuraStar Series
Страница 35: ...Chapter 2 Troubleshooting Procedures Page 2 2 DuraStar Series Stretcher Service Manual man268 NOTES ...
Страница 57: ...Chapter 3 Theory of Operation Page 3 2 DuraStar Series Stretcher Service Manual man268 NOTES ...
Страница 95: ...Chapter 5 Parts List Page 5 2 DuraStar Series Stretcher Service Manual man268 NOTES ...
Страница 97: ...Warranty Chapter 5 Parts List Page 5 4 DuraStar Series Stretcher Service Manual man268 NOTES ...
Страница 102: ...Recommended Spare Parts Chapter 5 Parts List DuraStar Series Stretcher Service Manual man268 Page 5 9 5 NOTES ...
Страница 137: ...Mattresses Chapter 5 Parts List Page 5 44 DuraStar Series Stretcher Service Manual man268 NOTES ...
Страница 139: ...Chapter 6 General Procedures Page 6 2 DuraStar Series Stretcher Service Manual man268 NOTES ...
Страница 165: ...7 9 Security Straps Chapter 7 Accessories Page 7 18 DuraStar Series Stretcher Service Manual man268 NOTES ...