EN
64/70
00503
-
new
Actual runtime (minutes)
R
00504
-
new
Set runtime (seconds)
RW
00505
-
new
Actual runtime (seconds)
R
00512
Display RCF or RPM mode in the display
Note:
If the radius (parameter 00620) is changed, the display is
automatically in RCF mode.
If set speed (parameter 00603) is changed the display is
automatically in RPM mode.
RW
00518
-
new
Actual program number.
Active program number (0..99) , which will be executed on next
start of centrifugation
or active program number (0..99) while running .
R
00519 -
new
Program
info
R
00520
-
new
Software lock command; If LOCK 5 is set, start of centrifugation
with the Start - key on the control panel is blocked.
RW
00521
-
new
Control command , start and stop centrifugation
START of the centrifugation is only possible if ALL of the following
conditions are TRUE:
- lid locking is closed.
- the hatch is closed (part 1) = the switch that detects
"lid is closed" is contacted,
- the hatch is closed (part 2) = the witch that detects
"hatch is closed" is contacted,
- the positioning mode is not active.
W
00522
-
new
Enable the active/current program block
W
00523
-
new
Program command , recall/activate or store/activate programs
W
00524 -
new
Positioning
command:
Lowbyte: 6-bit digit, which indicates the number of the target rotor
position ,
Range: 1 up to maximum (number of rotor positions ).
Highbyte: Even number of rotor positions,
2 up to 48, depending to the rotor.
RW
00526
-
new
Positioning and hatch command
W
00528
-
new
Positioning and hatch state
R
00537
-
new
centrifuge type and version
R
00563
-
new
Actual cycles of the actual rotor (High Word)
R
00564
-
new
Actual cycles of the actual rotor (Low Word)
R
00565
-
new
Preset cycles of the actual rotor (High Word)
R
00566
-
new
Preset cycles of the actual rotor (Low Word)
R
00567
-
new
Total cycles of the actual rotor (High Word)
R
00568
-
new
Total cycles of the actual rotor (Low Word)
R
00569
-
new
Number of centrifugation starts (High Word)
R
00570
-
new
Number of centrifugation starts (Low Word)
R