33/70
Parameter-No.: 00612
Run-down / braking
Gen. 1 + 2
Access:
RW
Structure:
MSD LSD MSD LSD
Lowbyte: Run-down parameter
Highbyte: Run-down parameter
Highbyte:
7
1= Level , 0= Time
6
5
4
3
2
1
0
7-bit digit MSB
Lowbyte:
7
6
5
4
3
2
1
0
8-bit digit LSB
1
st
possibility: run-down level
run-down level, if Highbyte -Bit 7- =1
range: 0 up to 9
2
nd
possibility: run-down time
run-down time, if Highbyte -Bit 7- =0
time in second
range: 1 up to 5 999 second ( 99 min : 59 sec )
Values out of minimum or maximum are automatically set to the minimum or maximum
value.
Minimum- / maximum-value: see parameter 00615 and 00616
Conversion from level to the corresponding time or reverse:
Here is a table necessary to allocate level and time to each centrifuge and each rotor.