EN
52/70
3.6.6
Recall and activate an already stored program
Example : Recall and activate program number 5
Please refer also to parameter 00523
3.7 Protocol
Examples
The following example shows a typically serial communication routine:
1. Start-up sequence (after centrifuge mains on during the beginning of serial communication
2. Cyclic sequence during standstill and positioning
3. Change the Nominal Values in stand still or during the run
4. The PC writes a program which will be stored in the control panel
(compatible with ROTANTA 46 RSC Robotic type 4815 / 4816 / 4817 )
5. The PC recalls and activates a program
(compatible with ROTANTA 46 RSC Robotic type 4815 / 4816 / 4817)
6. Command to open the hatch is sent and waiting for the execution of the command
7. Command for position 1 is sent and waiting for the execution of the command.
8. Command for position 4 is sent and waiting for the execution of the command.
9. Command to close the hatch is sent and waiting for the execution of the command
10. Prearrangement for a centrifugation:
Command for the recall of program 6 is sent and waiting for the execution of the command
11. Command to START the centrifuge is sent and waiting for the execution of the command.
12. Start
Note: The key-operated switch has to be in position LOCK 2 (right key position).
The centrifuge accept commands from PC only in position LOCK 2 .
STX = 02 hex , ETX = 03 hex , EOT= 04 hex , ENQ = 05 hex ,ACK = 06 hex , NAK = 15 hex
Abbreviation for column 1 :
E = ENQUIRY – telegram (value request) ,
S = SELECT – telegram (data-, or command-transmission).