5.10 Change the Nominal Values in stand still or during the run
Note: - With "Generation 2" it is not necessary to set LOCK 5,
but it will not cause a problem if LOCK 5 is set (to be confirm with "Generation 1"
- Multiple parameters could be modified by one procedure
but it will not cause a problem if only one parameter is modified (to be confirm with "Generation 1"
1. Write an new nominal parameter:
006XX = 00601, 00603, 00606, 00611, 00612, 00617, 00618 or 00620
PC
EOT
ADR
STX
0 0 6 x x = V V V V ETX
BCC
Cent
ADR ACK
PC
EOT
2. Next: SELECT: execute the change of the nominal value and still leave LOCK 5.
PC
EOT
ADR
STX
0 0 6 3 3 = 0 0 8 8 ETX
BCC
Cent
ADR ACK
PC
EOT
Is the nominal value "out of range", in SIOF (00685) bit 7 will be set and the centrifuge answers with:
Cent
ADR NAK
Then enquiry is necessary:
ENQUIRY to parameter 00685
PC
EOT
ADR
0 0 6 8 5 ENQ
Cent
ADR
STX
0 0 6 8 5 = V V V V ETX
BCC
PC
EOT
This Error bit is erased after call up of SIOF and the centrifuge can be input again by SELECT.
Note: If a program was used before, the centrifuge stores the modification under program 100 (corresponds to the
display indication: Program "----").
5.11 Recall (RCL) a program stored in the control panel, ready to start
Note: - With "Generation 2" it is not necessary to set LOCK 5,
but it will not cause a problem if LOCK 5 is set (to be confirm with "Generation 1"
Write Program-No. in parameter 00631 Highbyte , Bit 0...Bit 6 (xx) and set in Lowbyte, bit 2.
PC
EOT
ADR
STX
0 0 6 3 1 = x x 0 4 ETX
BCC
Cent
Program is shown in the display and program is active
Cent
Program will be registered in parameter 00630
Cent
Delete contents of 00631
Cent
Update of parameter 00634
Cent
ADR ACK
PC EOT
5.12 Change and store the values of a program
Note: - With "Generation 2" it is not necessary to set LOCK 5,
but it will not cause a problem if LOCK 5 is set (to be confirm with "Generation 1"
1. Write new nominal parameters:
006XX = 00601, 00603, 00606, 00611, 00612, 00617, 00618 or 00620
PC
EOT
ADR
STX
0 0 6 x x = V V V V ETX
BCC
Cent
ADR ACK
PC
EOT
2. Write Program-No. in 00631 Highbyte , Bit 0...Bit 6 (xx) and set in Lowbyte, bit 3 and bit 4.
PC
EOT
ADR
STX
0 0 6 3 1 = x x 1 8 ETX
BCC
3.Recall and activate program
PC
EOT
ADR
STX
0 0 6 3 1 = x x 0 4 ETX
BCC