Parameter-No.: 00473
Internal operating hours (Low Word)
Gen. 2
Access:
R
Structure:
MSD LSD MSD LSD
Lowbyte: Internal operating time
Highbyte: Internal operating time
Internal operating hours (Low Word)
Range: 0000H…FFFFH
The operating hours are stored as "long value" in seconds.
Calculation : Value in seconds = (High Word * 65536) + Low Word
Parameter-No.: 00474
Number of centrifugation starts
Gen. 2
Access:
R
Structure:
MSD LSD MSD LSD
Lowbyte: Number of starts
Highbyte: Number of starts
The value is the totally number of centrifugation runs, the range is:
0000H…FFFFH (0 – 656535)
2.10.1.3 Time
parameters
The time parameters of the “Generation 2” :
-
set time ( 00500, 00502 and 00504 )
-
actual time ( 00501, 00503 and 00505 )
are alternative to the parameters of the “Generation 1”
-
set time ( 00601 )
-
actual time ( 00602 ).
Note:
- The value range of the 005xx time parameters is up to 99 hours, 59 minutes and 59 seconds
- The value of the parameters 00601 and 00602 are restricted to 59999 seconds
(16 hours 39 minutes and 59 seconds).
Please refer also to parameter 00513 (Dual Timing Mode)
Parameter-No.: 00500
Set runtime (hours)
Gen. 2
Access:
RW
Structure:
MSD LSD MSD LSD
Lowbyte: Set runtime hours
Highbyte: Always 00
Input:
set runtime hours
Range:
0000H…0063H ( = 0d .. 99d) hours