52
This
oximeter measures functional saturation-oxygenated
hemoglobin expressed as a percentage of the hemoglobin that can
transport oxygen. It does not detect significant amounts of
dysfunctional hemoglobin, such as carboxyhemoglobin or
methemoglobin.
In contrast, hemoximeter such as the IL482 report fractional
saturation-oxygenated hemoglobin expressed as a percentage of
all measured hemoglobin, including measured dysfunctional
hemoglobins.
To compare functional saturation measurements to those from an
instrument that measures fractional saturation, fractional
measurements must be converted.
8.3
Measured
Versus Calculated
Saturation
When saturation is calculated from a blood gas partial pressure of
oxygen (PO
2
), the calculated value may differ from the SpO
2
measurement of a pulse oximeter. This usually occurs because the
calculated saturation was not appropriately corrected for the
effects of variables that shift the relationship between PO
2
and pH,
the partial pressure of carbon dioxide (PCO
2
), 2,3-DPG, and fetal
hemoglobin.
Содержание HS-Vet
Страница 1: ......
Страница 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Страница 98: ...97 Configuración del sistema Configuración de alarma ...
Страница 111: ...110 lectura manual debe corresponder con la lectura de la frecuencia del pulso que aparece en el monitor ...
Страница 119: ...118 Use un adhesivo para el sensor Use un nuevo sensor con adhesivo fresco Mantenga al animal inmóvil si es posible ...
Страница 136: ...135 pulsioxímetro ...
Страница 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Страница 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Страница 237: ...236 Alarm Setup ...
Страница 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...