35
Set the SpO
2
alarm limits as follows:
Choose
SpO
2
Hi Alarm
in the menu, press the
Function
button to enter setup. The
SpO
2
Hi Alarm
box will change from real line box to broken line box.
The adjustable range for upper limit of SpO
2
is from
“1 + the lower limit of SpO
2
” to 100. If the value of
SpO
2
Hi Alarm is set to less than 85, it will restore to
default value after the oximeter is turned on again.
Press the
Up Arrow
or
Down Arrow
button to
increase or decrease values.
Choose
SpO
2
Lo Alarm
in the menu, press the
Function
button to set up it. The
SpO
2
Lo Alarm
box
will change from real line box to broken line box. The
adjustable range for the lower limit of SpO
2
Alarm is
from 0 to “the upper limit of SpO
2
Alarm - 1”. If the
value of SpO2 Lo Alarm is set to less than 85, it will
restore to 85 after the oximeter is turned on again.
Press the
Up Arrow
or
Down Arrow
button to
increase or decrease values.
SpO
2
Hi Alarm
is always higher than
SpO
2
Lo Alarm
by at least 1%.
Press the
Function
button, confirm the alarm range
setup.
Press the
On/Off
button to exit the menu, and return to
measurement state.
3.3.10 PR Alarm Setup
The user can use
PR Hi Alarm
and
PR Lo Alarm
in menu to
adjust pulse rate alarm limits.
By default,
PR Hi Alarm
and
PR Lo Alarm
are 170 and 40
respectively.
Set the PR limits as follows:
Choose
PR Hi Alarm
in the menu, press
the
Function
button to enter setup. The
PR Hi Alarm
box changes
from real line to broken line. The adjustable range of
Содержание HS-Vet
Страница 1: ......
Страница 65: ...64 Appendix Ⅲ Record Table ID No Name Time SpO2 PR NOTE ...
Страница 98: ...97 Configuración del sistema Configuración de alarma ...
Страница 111: ...110 lectura manual debe corresponder con la lectura de la frecuencia del pulso que aparece en el monitor ...
Страница 119: ...118 Use un adhesivo para el sensor Use un nuevo sensor con adhesivo fresco Mantenga al animal inmóvil si es posible ...
Страница 136: ...135 pulsioxímetro ...
Страница 139: ...138 Apéndice III Tabla de registro Nº ID Nombre Hora SpO2 PR NOTA ...
Страница 204: ...203 Annexe Ⅲ Record Table No ID Nom Temps SpO2 PR REMARQUE ...
Страница 237: ...236 Alarm Setup ...
Страница 278: ...277 Anhang III Tabelle für Aufzeichnungen ID Nr Bezeichnung Zeit SpO2 PF HINWEIS ...